Specifications |
|
ASTM D910, D4814, D6227, EN 228 |
Available Test Methods |
|
ASTM D5188, D5191, D6377, D6378, D6897
excellent correlation to D323, D4953, D5190, D5482, D1267
EN13016-1 & EN13016-2, IP394, IP409, IP481
JIS K2258-2, SHT 0769, GOST 52340;
customized measuring programs |
Vapor-Liquid ratio |
|
Variable from 0.02/1 to 100/1, single and multi-point & V/L curve measurements |
Temperature range |
|
0°C to 120°C (32 to 248 °F); Peltier technology, no external cooling necessary |
Temperature Stability |
|
0.01°C (0.014°F) |
Temperature profiles |
|
Single temperature, stepped or ramped |
Pressure range |
|
EV01 ERAVAP: 0 to 1000 kPa (0 to 145 psi) EV02 ERAVAP LPG: 0 to 2000 kPa (0 to 290 psi) |
Pressure resolution |
|
0.01 kPa; Accuracy: 0.2 kPa (0.028 psi) |
Pressure units (selectable) |
|
hPa, kPa, psi, mmHg, bar, mbar, at, atm |
Precision |
|
Repeatability: r = 0.3 kPa, Reproducibility: R = 0.7 kPa |
Sample introduction |
|
Automated via built in piston, no external vacuum pump required |
Measuring time |
|
5 minutes for a standard measurement |
Shaker motor |
|
Built-in shaker motor for fast crude oil & T(V/L) measurements |
D5191 & D5188 Speed test |
|
VP & T(V/L=20) measurement of gasoline at the push-of-a-button |
Measuring chamber |
|
Nickel plated aluminum chamber |
Sample volume |
|
1 ml (2.2 ml per rinsing cycle) |
Cleaning |
|
By next sample or solvent |
Display |
|
Industry proven 7’’ color touch screen |
Remote Control |
|
Remote service capability via Ethernet interface |
Interfaces |
|
Built-in PC with Ethernet, USB and RS232 interfaces; direct LIMS connectivity and output to printer or PC |
Auto sampler |
|
Optional 10 position auto sampler – directly attached! |
Quality-Check tracking |
|
Statistical quality assurance & graphical control charting (ASTM D6299) |
Result database |
|
> 100,000 detailed test reports stored in internal memory: DVPE, RVPE, PABS, PGAS, PTOT, ASVP & AVP, T(V/L), VLI, TVP |
Operators |
|
Unlimited operator names memorized |
Hazy mark |
|
Results can be marked for hazy appearance |
Sample container size record |
|
100 ml, 250 ml, 1 l |
Alarm tracking |
|
All messages are stored in the database together with the result |
Automated calibration routine |
|
10-point probe offset correction table, calibration history tracking, self-diagnostics |
Power requirements |
|
Auto switching 85-264 V AC, 47-63 Hz, 100 W (multi-voltage power supply)
Field application: 12 V/9A DC (vehicle battery) |
Dimensions / Weight |
|
22 x 32 x 28 cm (8.7 x 12.6 x 11 in) / 8 kg (17.6 lb) |