Test standards |
|
ASTM D1250, D1475, D4052, D4806, D5002, D5798, D5931, ISO 12185, ISO 15212 |
Measuring method |
|
Resonant frequency oscillation |
Measurement Range |
|
0 ~ 3 g/cm3 |
Temperature Range |
|
0 ~ 96°C (32 ~ 204.8°F) |
Accuracy |
|
Density:
Models DA-650 DA-650B: ±2×10-5 g/cm3 (0.00002 g/cm3)
Models DA-645 DA-645B: ±5×10-5 g/cm3 (0.00005 g/cm3)
Models DA-640 DA-640B: ±1×10-4 g/cm3 (0.0001 g/cm3)
(Calibration with air and water required.)
Temp:
Models DA-650 DA-650B: ±0.02 °C (±0.04 °F)
Models DA-645 DA-645B: ±0.03 °C (±0.05 °F)
Models DA-640 DA-640B: ±0.05 °C (±0.09 °F) |
Repeatability*2 |
|
Density SD 5×10-6 g/cm3, SD 1×10-5 g/cm3, SD 5×10-5 g/cm3 |
Minimum Sample Required |
|
1) Approx. 1 mL (Syringe)
2) Approx. 2 mL (Pump) |
Measurement Time |
|
1) 1 to 4 mins (Manual)
2) 2 to 10 mins (Auto) |
Display |
|
1) 5.7-inch colour TFT LCD; 640 x 480
2) Shows density, specific gravity, oscillation frequency, temperature, concentration and other messages |
Resolution |
|
Density & Specific Gravity: 0.000001
Temperature: 0.001 |
Auto calibration |
|
Density of water at each temperature, required for calibration of measuring cell, is pre-installed. |
Sampling by |
|
Syringe-type hand pump |
Viscosity Correction |
|
Models
DA-650 – Yes
DA-650B – No
DA-645 D – Yes
A-645B – No
DA-640 – Yes
DA-640B – No |
Sampling |
|
1) Manual by syringe
2) Auto by sampling pump*3 |
Method |
|
Saves up to 100 different methods in built-in memory |
Stability |
|
Four modes of stability according to measurement accuracy and time |
Density Auto Correction |
|
1) Saves conversion table or formula at your desired temperatures according to your samples.
2) Temperature conversion table preprogrammed according to ASTM standard for petroleum, petroleum products and lubricating oils |
Auto Conversion |
|
1) Between concentration and density
2) Between temperature and density |
Statistics |
|
1) Auto or manual calculation of mean value, SD and coefficient
2) Recalculation, data deletion |
Interfaces |
|
1) LAN: x 1; Personal computer (PC)
2) USB 1.1: x 2; USB flash drive, keyboard, barcode reader, Epson inkjet printer *4
3) Software: x 2; Dot Matrix Printer, Auto Clean and Sampling Unit, Multiple Sample Changer |
Options |
|
1) Printer: Thermal Printer, Dot Matrix Printer
2) Sampling Unit, Changer: DCU-551N/H, CHD-502N/H/C
3) Software: SOFT-CAP (Data Acquisition Software) |
Data I/O |
|
1) USB flash drive as data storage medium
2) Application Notes provided in USB flash drive |
Wetted Materials |
|
PTFE, borosilicate glass, SUS304 |
Ambient Conditions |
|
1) Temperature: 5 ~ 35 °C (41 ~ 95 °F)
2) Humidity: 85%RH or below (No condensation allowed.) |
Power Supply |
|
AC 100 ~ 240V; 50/60Hz (Comes with AC adapter.) |
Power Consumption |
|
40W (max. 120W, min. 20W) |
Dimensions |
|
320 (W) x 365 (D) x 250 (H) mm (12.6 (W) x 14.4 (D) x 9.8 (H) inches) |
Weight |
|
18 kg (39.7 lbs) |
Export Packing in Double Carton Box |
|
G/W 21 kg (46.3 lbs); 540 (W) 480 (D) 460 (H) mm (21.3 (W) 18.9 (D) 18.1 (H) inches) (May vary in some cases.) |