FREQUENTLY ASKED QUESTIONS
Errors
Getting Error E004 (Eraflash)
This error is because the cup of the oven plate is not clean. Physically inspect the sample cup. Ensure no dents, scratches are on the cup. Clean the oven plate gently with using isopropanol. If still getting the error, send the machine for the repair.
Getting Error E004 (Eravap)
This error occurs when pressure exceeds 1000KPa for regular and 2000KPa for LPG unit. This shows pressure exceeds the upper limit. Don’t run the machine. Send the machine for the repair. It might be an issue either with pressure transducer or inlet/outlet valves.
How To’s
How to perform intensive valve cleaning?
Go to the service menu
Under sub tab service
Press intensive valve cleaning
machine will ask to connect inlet and outlet tubing and put both of them in Toluene
After some time machine will ask to remove the inlet tube
How to refill the piston oil?
Pull out the metal casing by pressing the two push buttons on the back of the machine or unscrew the 2 screws
Check the oil level
If the oil level is below the mark
Open the oil level container screw
If the oil level is low fill the oil up to the mark
then go to the Service menu
under sub tab service Press start for refill oil
Once a message appears to close the screw. Close the screw and press finished on the screen.
How to reset the filter counter?
Before resetting the filter first inspect the condition
Open the inlet luer
If particles are present on the filter
Clean it or replace with the new one
Mesh filters are the first safety for the valves and the system
Go to the Setting
Under sub tab Filter
Press Reset
How to check the live pressure of the unit?
Eravap:
Go to the Service menu
Open inlet/outlet valves
On the right top corner the machine will show the live pressure
Eraflash:
Go to the Service menu
On the right top corner live pressure reading
How to clean the ignitor?
Go to the service
Under sub tab Test there is an option Ignition
Press Clean
Just press one or two times
How to perform the air saturation?
Returns
Do I need a return authorization number to send my unit in?
Yes, please contact the service department or fill out an online form.
What is your address?
Ayalytical Instruments
1022 Hercules Ave
Houston, TX 77058
Ayalytical Instruments
2787 W. Fulton St.
Chicago, IL 60612
Contact Us
Contact Us and Ask a New Question