Advanced flash point tester, automated for 10 samples for highest efficiency.
- Safe and quick measurements
- Small sample volume required for up to 10 automated measurements
- Wide temperature range for diverse applications, ability of connecting to High Temperature Module (HTM) extension to increase the temperature range up to 420°C (788°F)
- Fast results with an easy-to-use touchscreen interface
- Portable and robust for field use
Key Features

ASTM
CompliantFully compliant with ASTM D6450, D7094, and D5808, providing both standard method and refined methodology

QuickQuick: Innovative PBT – Peltier Boost Technology TM, a patented eralytics innovation, allows for previously unmatched heating and cooling rates. Paired with a conveniently small sample size, these analyzers allow turnaround times that are significantly shorter than those offered by all other instruments available on the market

Optional
AutosamplerOptional Autosampler: A 10-position autosampler can be added on to maximize testing volume and time

Temperature ControlTemperature Control: Standalone control range of 0 °C–200 °C (32 °F–390 °F) with no external cooling required. Down to -25 °C (-13 °F) with water cooling or chiller, and up to 420 °C (788 °F) with the optional High Temperature Module (EF01-HTM)

Touch
ScreenTouch Screen: All eralytics analyzer results are displayed immediately after the test on their large and user-friendly color touch screen

Programmable MethodsProgrammable Methods: Operators can create and save their own test methods. Any method saved to memory can be recalled and run directly from the instrument

Easy
CleanEasy Clean: Cleaning after measurements often took time and involved maintenance activities or replacing sensors or other elements of the chamber assembly. Innovative CPT – Contamination Prevention Technology TM, a combination of an advanced electrode protection and self-cleaning ignition system, reduces cleaning and maintenance to an absolute minimum

Measures
All FuelsMeasures All Fuels: Specifies fuels, analyzes fresh and used lubricating oils, and tests bitumen. The small sample volume also renders it ideal for more expensive samples such as flavors and fragrances, cosmetics, paints and varnishes, or even hazardous waste

Safety
AlarmsSafety Alarms: All alarm messages are stored in the database together with the result. The instrument automatically shuts off and the problem is reported by buzzer and display
Videos


Built-in 10-port autosampler
Ten samples.
One unattended run.
Introducing the ERAFLASH S10: augmented ultra-wide range flash point tester maximized for high-volume autosampling
The ERAFLASH S10 is a compact, safe, and reliable flash point tester covering an incredibly wide temperature range with a built-in 10-port autosampler. With a continuously closed cup design paired with unparalleled thermoregulation using Peltier elements, the ERAFLASH S10 ensures maximized safety and efficiency while subsequently testing up to ten samples in one run.
High-volume autosampling
Load the sequence.
Let S10 keep testing.
1–2 mL
Sample per test
Continuously closed cup design
With the distinct capability to incorporate both a high temperature and a low temperature module, this singular analyzer can cover a temperature range from −25 °C to 420 °C with only 1–2 mL of sample per test. To complete a test, the device produces an electric arc in a closed chamber to determine at which temperature the arc ignites the sample vapor, which is measured by the pressure change observed in ignition. The notable feature of the ERAFLASH S10 is the built-in autosampler, making it the ideal solution for unattended flash point testing.
Built for unattended flash point testing
Compliance / Correlation
Standards-ready.
High-volume capable.
ASTM D6450 · ASTM D7094
This device is fully compliant with ASTM D6450 and D7094 with a diverse array of correlation methods, making it the ideal solution for almost any flash point determination in both laboratory and on-the-go environments.
- 01
Temperature Range
−25 °C to 420 °C (−13 °F to 788 °F). Stand-alone range is 0 °C to 200 °C; the low end requires external cooling and the high end the optional High Temperature Module.
- 02
Flash Point + Flash/No Flash
Flash point corrected to sea level is displayed in units of temperature. In correlation tests: Flash versus No Flash determinations.
- 03
Pressure Change + Temperature
Largest pressure difference during the flash test. Live, expected, and corrected temperatures.
- 04
Unattended Throughput
Ten loaded samples in approximately 45 minutes, depending on temperature range.
Applications of the ERAFLASH S10
- Safety characteristic determination
- Transportation regulations compliance
- Appropriate storage requirements
- Aids in fire protection and investigation
Specs
| Standard Test Methods | ASTM D975, ASTM D1655, ASTM D2880, ASTM D3699, ASTM D6450, ASTM D7094, ASTM D7467, ASTM D7566, ASTM D8606, IP 620, ISO 24966, and SH/T 0768 | |
| Fuel Specifications (ASTM D7094) | ASTM D396, ASTM D975, ASTM D2880, ASTM D3699, ASTM D7467 | |
| Approved By | NATO, RCRA, US DOT, US NAVY | |
| Special Test Methods | Fast screening test method SCAN for unknown samples, Flash / No Flash tests, freely programmable test methods | |
| Correlation Test Methods | Pensky Martens Closed Cup: ASTM D93; EN ISO 2719; DIN 51758; IP34; JIS K2265 TAG Closed Cup: ASTM D56 Abel Pensky Closed Cup: ISO 13736; IP170 Small Scale Closed Cup: ASTM D3828; EN ISO 3679, EN ISO 3680 |
|
| Sample Handling | 10 position autosampler | |
| PBT – Peltier Boost Technology | High speed heating & cooling -25 °C to 420 °C (-13 °F to 788 °C) with a single analyzer |
|
| CPT – Contamination Prevention Technology | Advanced electrode protection and self-cleaning ignition system to minimize cleaning and maintenance | |
| Combustion Graphics | Display of combustion characteristics for contamination analysis | |
| QuickCal | In-situ calibration with eralytics unique calibration cups without the need of dismantling the instrument | |
| Fuel Dilution Program | Automatic fuel dilution measurement for the analysis of used engine oils | |
| Temperature Range | 0 °C to 200 °C (32 °F to 390 °F) stand alone down to -25 °C (-13 °F) with external cooling up to 420 °C (788 °F) with optional High Temperature Module |
|
| Temperature Stability | 0.1 °C (0.2 °F) | |
| Unattended Sample Troughput | 10 loaded samples within approximately 45 minutes (depending on temperature range) | |
| Sample Volume | 1 mL (ASTM D6450), 2 mL (ASTM D7094), ISO 24966 | |
| Sample Cups | Nickle plated aluminium, stainless steel cups available upon request | |
| Display | Industry proven 8.4″ color touchscreen | |
| Languages | English, German, French, Spanish, Russian, Chinese Additional langages available upon request |
|
| Interfaces | Built-in PC with Ethernet, USB and RS232 interfaces Direct LIMS connectivity and output to printer or PC Optional input by external keyboard, mouse and barcode / QR-code reader |
|
| Remote Control | Remote service capability via Ethernet interface | |
| PC Software | ERASOFT RCS – remote control Windows software for multi-instrument remote control, convenient data transfer and result analysis | |
| Result Database | Over 100 000 detailed test reports stored in internal memory | |
| Alarm Tracking | All alarm messages are stored in the database together with the results | |
| Interfaces | Wifi via external USB dongle | |
| Operating Conditions | Temperature range 10°C to 40°C Humidity up to 90% RH, non condensing |
|
| Power Requirements | Auto-switching 85-264 V AC, 47-63 Hz, max. 150 W (multi-voltage power supply) Field application: 12 V DC (vehicle battery) adapter available |
|
| Dimensions / Weight | (W x D x H) = 29.5 x 36.9 x 45.4 cm (11.6 x 14.5 x 17.9 in) / 18.1 kg (39.9 lb) |