ERADENS XS
Battery-operated Ultra-compact highly accurate density meter for on-site density measurement.
- High accuracy and repeatability
- Ultra-Lightweight and easy to carry
- Quick measurements and simple interface
- Battery-Operated with Bluetooth Connectivity
- Meets different standards
Key Features
Bluetooth-connected / Field-ready
Density
goes mobile.

Introducing the ERADENS XS: ultra-portable high-precision density measurement
The ERADENS XS is a state-of-the-art, ultra-portable density meter designed for both laboratory and field operations. Its compact design and oscillating metal U-tube technology deliver accurate density measurements across petroleum products, chemicals, and beverages. Compliant with ASTM D4052, ASTM D5002, and ISO 12185, ERADENS XS supports precise, fast, and reliable results.
One sample / Connected results
Measure here.
Review anywhere.
01
1 ml
Sample
02
15 sec
Result
03
App +
Share
At just 11 × 16 × 14 cm (4.4 × 6.3 × 5.6 in) and 1.5 kg (3.3 lb), ERADENS XS is exceptionally portable and can operate on battery power. Results arrive within 15 seconds of temperature equilibrium, with a density resolution of 0.00001 g/cm³ and accuracy up to 0.00005 g/cm³.
Battery · Bluetooth · iOS + Android app
Independent or integrated
One device.
Two ways to work.
- Stand-alone
Independent density testing
Operate the analyzer on its own, anywhere the sample is.
- Connected
Combined measurements with an external analyzer
Link to an eralytics analyzer over micro-USB for combined results.
ERADENS XS offers Bluetooth connectivity and control through a dedicated iOS and Android app for seamless operation, data storage, and result sharing. It can connect to an external analyzer for combined measurements or operate independently. Built-in bubble detection verifies accurate sample filling by preventing air bubbles from compromising results.
Applications of the ERADENS XS
- Precise density measurement in remote or field locations
- Quality control in the beverage industry for alcohol and sugar content
- Research and development requiring reliable density data
ASTM D4052 · ASTM D5002 · ISO 12185
Specs
| Standard Test Methods | ASTM D4052, D5002, D5931, D7777, ISO 12185 |
| Density Range | 0 – 3 g/cm3 |
| Temperature Range | 10 °C to 60 °C (50 °F to 140 °F) |
| Density Accuracy | 0.00005 g/cm3 |
| Density Repeatability (SD) | 0.00001 g/cm3 (according to ISO 5725) |
| Density Resolution | 0.00001 g/cm3 |
| Temperature Accuracy | 0.02 °C (0.04 °F) |
| Measurement Time | 15 sec (after temperature equilibration) |
| Minimum Sample Volume | 2 mL |
| Oscillating Tube | Hastelloy (Patent: US 11,630,044) |
| Bubble Detection | Yes |
| Viscosity Correction | Yes – full range correction (patented: AT523471, A1) |
| Operation | Mobile instrument operation with the eralytics app ‘eradens xs’ (available in all App stores) |
| Tables & Functions | API, sugar, alcohol, etc. |
| Interfaces | Micro-USB: Communication to an eralytics analyzer Bluetooth®: Communication to mobile device (mobile phone, tablet) |
| Power Requirements | 110 – 240 VAC, 50 – 60 Hz, 120 W |
| Battery | 8 h at 15 °C measuring temperature |
| Operating Conditions | 5 °C – 45 °C up to 90 % rel. humidity (non-condensing) |
| Dimensions & Weight | W x D x H = 11 x 13 x 16 cm (4.3 x 5.1 x 6.2 in) / 1.4 kg (3.2 lb) With battery: W x D x H = 11 x 13 x 21 cm (4.3 x 5.1 x 8.3 in) / 2.1 kg (4.5 lb) |