Automated Pour/Cloud Point Tester

Advanced tester for measuring pour and cloud points in petroleum products.

Key Features:

  • Precise and reliable measurements.
  • Automated operation for ease of use.
  • User-friendly interface with clear readouts.
  • Robust design for consistent performance.
  • Compliant with ASTM and ISO standards.
Methods:
ASTM, D97, D2500, D6749, D7683, ISO, 3016, 3015

Login or Register to buy or

[request_quote_button]

TANAKAs MPC series cloud point analyzer has been designed for automatic determination of POUR POINT (PP) and CLOUD POINT (CP) with small specimen size and shorter test cycle time while securing better test precision than the conventional manual methods.

PP measurement on this cloud and pour point apparatus is by AIR PRESSURE METHOD (ASTM D6749), and CP measurement is by SMALL TEST JAR METHOD (ASTM D7683).

The epoch-making automatic PP test method yields 1C test resolution, while the new CP method yields 0.1C resolution.

Features

features__astm compliantfeatures__simple operatonfeatures__quickfeatures__high precisionfeatures_compact

Models

Models Positions
MPC-102A Air-Cooled
MPC-102L Liquid-Cooled

Specs

CONFORMING STANDARDS PP: ASTM D6749, ASTM D97, ISO 3016 / CP: ASTM D7683, ASTM D2500, ISO 3015
TYPE Mini Pour Point(PP) and Cloud Point(CP) tester with sequential CP and PP measuring capability. Sample cooling and pre-heating by TED.
SPECIMEN VOLUME  4.5ml
MEASURING RANGE (typical*) 1. MPC-102L(Liquid cooled model): +51oC to -40oC with tap water of 20oC +51oC to -65oC with cooling liquid of -35oC 2.MPC-102A(Air cooled model): +51oC to -30oC (in 25oC ambient) *:Sample viscosity, etc. affects on lowest tempera- ture of the measuring range.
MEASUREMENT MODES Selectable from various modes.
1.CP mode (0.1 or 1.0 oC, selectable)
2.PP modes: Programmed by the user. Programmable parameters are: *Amount of applied air pressure for PP detection, to accommodate different sample types: L(low) for diesel fuels, H(high) for lube oils, VH(very high) and UH (ultra high) for residual fuels and similar samples. *Testing intervals: 1.0oC, 2.5oC, or 3.0oC (In total, 4×3=12 modes for PP.)
3.CP/PP modes: CP is determined and then PP. PP detection is programmable by the user with the same parameters as PP modes’. (12 modes in total.)
SAMPLE AUTOMATIC PRE-HEATING Automatic preheating at either +45oC or EPP+10oC. (EPP=Expected Pour Point)
DISPLAY Test parameters, EPP, bath temperature, sample temperature, PP, and CP displayed on VFD. Temperatures displayed in 0.1oC increments.
EPP SETTING EPP(Expected Pour Point) needs to be set prior to test Starts.
SPECIMEN CUP Cylindrical glass test jar with 4.5ml sample volume.
SENSORS Compound type sensor assembly for PP and CP. PP detected by air pressure method(patented). CP detected photo-electrically. PT100 temp. sensors.
SAMPLE COOLING RATE As standard, 4oC/min. till EPP+40oC, and 1oC/min. thereafter. Cooling profile is programmable
SAFETY SHUTDOWN As hot side of TED reaches 60oC while preheating, warning buzzer beeps and heating stops.
DATA OUTPUT RS-232C 1 channel (for PC or Optional Printer)
DATA STRAGE Last 50 test data are stored in RAM
POWER REQUIREMENTS 100, 120, 220, or 240VAC 0.5kW
DIMENSIONS AND WEIGHT 230mmWx480mmDx385mmH
20kg

Downloads

AYALYTICAL EXCLUSIVE SERVICES

ISO OEM 17025 Certified

ISO 17025

Our Service Center has been ISO Accredited for Competence of Calibration.

AYALYTICAL EXCLUSIVE SERVICES  

1-3 Days Service Turn Around

 

We are please to offer our 1-3-day service turn around from our Ayalytical Service department.

AYALYTICAL EXCLUSIVE SERVICES

Application Support

 

We are pleased to offer our expertise in various disciplines including Physical Properties, Elemental and Compositional.

AYALYTICAL EXCLUSIVE SERVICES

Dedicated Support Lines

 

We are pleased to offer our On-Site trainings and service calls for our Ayalytical partners.