ERAVAP -DENSITY

Vapor pressure measurement with integrated density measurement. 

Key Features:

  • Simultaneous vapor pressure and density analysis
  • Fast and reliable results
  • Compact and portable design
  • Easy-to-use touchscreen interface
  • Complies with different standards
Methods:
ASTM, D4052, D5191, D6378, D5188

Login or Register to buy or

[request_quote_button]

Introducing ERAVAP-DENS: Vapor Pressure Testing with Built-In Density Analysis

Meet ERAVAP-DENS — the next evolution in vapor pressure testing, now with an integrated density module. Combining the precision of the ERAVAP with advanced density measurement, ERAVAP-DENS is the all-in-one solution for RVP and DVPE testing across a wide range of products, including fuels, solvents, fragrances, and aqueous solutions.

Designed with simplicity and efficiency in mind, ERAVAP-DENS delivers precise volatility and density data in a compact, fully automated device. Its rugged aluminum casing and intuitive interface make it ideal for both lab and field applications. With an optional 10-position autosampler, high-volume testing becomes effortless while retaining the mobility you expect.

Precision Meets Versatility

ERAVAP-DENS sets the standard for comprehensive testing. From gasoline performance analysis to emissions compliance, it provides the accuracy you need to optimize operations and ensure regulatory compliance. The added density module incorporates ASTM and ISO standards, making it perfect for R&D, QC, and beyond.

This device is fully compliant with ASTM D5188, D5191, D6378, and EN 13016 with an added density module incorporating ASTM D4052 and ISO 12185 making it the ideal unit for R&D, QC, and many more applications using a single analyzer.

Applications of the ERAVAP DENS Include:
Gasoline performance analysis
Emission regulations compliance
Seasonal vapor lock avoidance
Nature of refinery feedstocks

Key Features:

All-in-one testing for vapor pressure and density, providing essential data for a variety of applications.
Compact and mobile design, with a simple interface for minimal operator interaction.
Advanced measurements, including volatility profiles, pressure distribution, and vaporization tendencies.
Backed by a 1-year warranty with an optional extension, ERAVAP-DENS delivers the precision and reliability needed for today’s demanding environments.

Explore more: Discover the ERAVAP ONLINE or ERAFLASH, or request a personalized quote today.

Features

features__astm compliantERAVAP is fully compliant with ASTM D1267, ASTM D323, ASTM D4814, ASTM D4953, ASTM D5188, ASTM D5190, ASTM D5191, ASTM D6377, ASTM D6378, ASTM D6897, ASTM D910, providing both standard method and refined methodology.
features__simple operatonAll eralytics analyzers use an intuitive menu structure that is easy to navigate. Samples need no prior preparation and measurements are done fully automated.
A direct out-of-the-box connection to LIMS (Laboratory Information Management System) is established with a standard Ethernet cable.No extra software or PC are necessary.
features__footprintThe small footprint of the eralytics instruments makes them the first choice for use in the lab and on-site.
features__portableThe Eralytics units are all lightweight and built with an integrated carrying handle, they are versatile and ideally suited for mobile applications as well.
features__touch screenAll Eralytics analyzer results are displayed immediately after the test on their large and user-friendly color touch screen.
features__high precisionThe outstanding precision of the ERAVAP is the result of combining the innovative measurement cell design with a choice of cutting-edge instrument parts. Among those count a self-lubricating piston drive, a built-in shaker motor and state-of-the-art solenoid valves that prevent cross contamination between samples. The ERAVAP instrument clearly outperforms the standard methods in terms of repeatability and reproducibility. (r = 0.3 kPa, R = 0.7 kPa).

Accessories

The following accessories are recommended for purchase in sequential order (see eStore):

1st Year: No spare parts are required
2nd Year: Sampling tube with Luer connector
Sampling tube 0,5m with Luer connector
Teflon o-ring for EV01-A002a Mesh inlet filter 80µ
Outlet tube
Kalrez O-ring for pressure sensor (Kalrez 12.42×1.78)
O-ring small for measuring cell cover (Viton 6×1.5)
Remaining Parts:

Specs

Specifications ASTM D910, D4814, D6227, EN 228
Available Test Methods ASTM D5188, D5191, D6377, D6378, D6897
excellent correlation to D323, D4953, D5190, D5482, D1267
EN13016-1 & EN13016-2, IP394, IP409, IP481
JIS K2258-2, SHT 0769, GOST 52340;
customized measuring programs
Vapor-Liquid ratio Variable from 0.02/1 to 100/1, single and multi-point & V/L curve measurements
Temperature range 0°C to 120°C (32 to 248 °F); Peltier technology, no external cooling necessary
Temperature Stability 0.01°C (0.014°F)
Temperature profiles Single temperature, stepped or ramped
Pressure range EV01 ERAVAP: 0 to 1000 kPa (0 to 145 psi) EV02 ERAVAP LPG: 0 to 2000 kPa (0 to 290 psi)
Pressure resolution 0.01 kPa; Accuracy: 0.2 kPa (0.028 psi)
Pressure units (selectable) hPa, kPa, psi, mmHg, bar, mbar, at, atm
Precision Repeatability: r = 0.3 kPa, Reproducibility: R = 0.7 kPa
Sample introduction Automated via built in piston, no external vacuum pump required
Shaker motor Built-in shaker motor for fast crude oil & T(V/L) measurements
D5191 & D5188 Speed test VP & T(V/L=20) measurement of gasoline at the push-of-a-button
Measuring chamber Nickel plated aluminum chamber
Sample volume 1 ml (2.2 ml per rinsing cycle)
Cleaning By next sample or solvent
Display Industry proven 7’’ color touch screen
Remote Control Remote service capability via Ethernet interface
Interfaces Built-in PC with Ethernet, USB and RS232 interfaces; direct LIMS connectivity and output to printer or PC
Auto sampler Optional 10 position auto sampler – directly attached!
Quality-Check tracking Statistical quality assurance & graphical control charting (ASTM D6299)
Result database > 100,000 detailed test reports stored in internal memory: DVPE, RVPE, PABS, PGAS, PTOT, ASVP & AVP, T(V/L), VLI, TVP
Operators Unlimited operator names memorized
Hazy mark Results can be marked for hazy appearance
Sample container size record 100 ml, 250 ml, 1 l
Alarm tracking All messages are stored in the database together with the result
Automated calibration routine 10-point probe offset correction table, calibration history tracking, self-diagnostics
Power requirements Auto switching 85-264 V AC, 47-63 Hz, 100 W (multi-voltage power supply)
Field application: 12 V/9A DC (vehicle battery)
Dimensions / Weight 22 x 32 x 28 cm (8.7 x 12.6 x 11 in) / 8 kg (17.6 lb)

Downloads

Download our brochure for more details

Videos

Configuration

ASTM D5191 with D4052

AYA-83-22243 Sampling tube (FEP) 0,4m with Luer connector
AYA-82-22237 Mesh inlet filter 80µ (5 pcs, 1 pc inside Sample inlet on instrument)
AYA-82-22084 Outlet tube
AYA-82-22085 Waste container
AYA-82-22086 Power cable

ASTM D6378 with D4052

AYA-83-22243 Sampling tube (FEP) 0,4m with Luer connector
AYA-82-22237 Mesh inlet filter 80µ (5 pcs, 1 pc inside Sample inlet on instrument)
AYA-82-22084 Outlet tube
AYA-82-22085 Waste container
AYA-82-22086 Power cable

AYALYTICAL EXCLUSIVE SERVICES

ISO OEM 17025 Certified

Our Service Center has been ISO Accredited for Competence of Calibration.

AYALYTICAL EXCLUSIVE SERVICES  

1-3 Days Service Turn Around

 

We are please to offer our 1-3-day service turn around from our Ayalytical Service department.

AYALYTICAL EXCLUSIVE SERVICES

Application Support

 

We are pleased to offer our expertise in various disciplines including Physical Properties, Elemental and Compositional.

AYALYTICAL EXCLUSIVE SERVICES

Dedicated Support Lines

 

We are pleased to offer our On-Site trainings and service calls for our Ayalytical partners.