ERACHECK PRO

CFC-free oil-in-water testing with unrivaled precision.
Methods:
EPA 1664A, EPA 9071B, ASTM D3921, ASTM D4281, ASTM D7066, ASTM D7678, DIN 38409-H18, EN 14039, ISO ISO9377-1,

								
								
								
								
								
								
								
								
								
								
								
								EPA 1664A, EPA 9071B, 
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								
								ASTM D3921, ASTM D4281, ASTM D7066, ASTM D7678, 
								
								
								
								
								
								DIN 38409-H18, 
								
								EN 14039, 
								
								
								
								
								
								
								
								ISO ISO9377-1, 
								
								
								
								
								
								
								
								

Durable. Dependable. Design you can trust.

ERACHECK PRO is the industry-leading measurement device for CFC-free and reliable oil-in-water or oil-in-soil testing. The only fully ASTM D7678 compliant analyzer on the market, it combines all benefits of the withdrawn IR method ASTM D3921 with the advantages of CFC-free solvents. It measures sub-ppm concentrations of Total Petroleum Hydrocarbons (TPH) and Total Oil and Grease (TOG) in water and soil.

Accurate enough for lab work, while compact and robust for field applications, the ERACHECK utilizes the low-cost solvent cyclohexane, ensuring a safe, simple and environmentally friendly extraction procedure.

ERACHECK’s broad appeal can be attributed to its innovative QCL-IR (Quantum Cascade Laser Infra-Red) technology, which distinctly measures total oil and grease in water. Developed by the QCL specialists at QuantaRed, QCL-IR performs measurement across a wide range (0.5 to 2,000 ppm), a unit precision that exceeds ASTM D7678 standards and outperforms all other infrared techniques.

For more information about the QCL-IR oil and grease analyzer, please visit www.quantared.com.

To request a quote for the ERACHECK PRO, please click here.

features__astm compliantfeatures__simple operatonfeatures_direct_lims_communicationfeatures__footprintfeatures__portablefeatures__touch screen

The following accessories are recommended for purchase in sequential order (see eStore):

1st Year:
No spare parts are required
2nd Year:
Sampling tube with luer connector for ERACHECK
Metal filter 16 µ (5 pcs)
Mesh inlet filter 80µ (5 pcs)
Teflon o-ring for EV01-A002a Mesh inlet filter 80µ
Outlet tube
Protective film for touch display (5 pcs)
Stylus for data entry via touch display
Glas syringe 20 ml
Micro-Syringe (250 µl)
Volumetric flask (50 ml) with glass stopper
Remaining Parts:
Available test methods  

ASTM D7678-11

Applications   TPH (Total Petroleum Hydrocarbons) and TOG (Total Oil and Grease) measurements in water an soil
Excellent correlation to   Infrared Spectroscopy: ASTM D3921, ASTM 7066, IP426, DIN 38409-H18, OSPAR IR method
Gas chromatography: ISO 9377-2, ISO 9377-2 (mod) OSPAR, ISO 16703, EN 14039 NEN 5733, MADEP-EPH DIN 38409-H53
Gravimetry: EPA 1664B, EPA 9071B, ASTM D4281, EN 14345
Extraction Solvents (CFC-free)   Cyclohexane (recommended)
Cyclopentane (ERACHECK PRO only)
Extraction method   External liquid-liquid (solid-liquid) extraction
Sample Volume   Typically 900 mL water and 50 mL cyclohexane according to ASTM D7678
Sample Celan-up (TPH)   Simplified removal of polar substances over attachable Na2SO4 / Florisil® cartridges
Measurement Time   2 min (including background measurement)
Measurement Range   oil-in-water testing
up to 2 000 mg/L (ERACHECK PRO)
up to 200 mg/L (ERACHECK) 
oil-in-soil testing
up to 72 000 mg/kg (ERACHECK PRO)
up to 7 200 mg/kg (ERACHECK)
Limit of Detection   oil-in-water testing
0.2 mg/L (900mL water : 50 mL solvent, ERACHECK PRO)
<0.1 mg/L (900mL water : 20 mL solvent, ERACHECK PRO)
1.5 mg/L (900mL water : 50 mL solvent, ERACHECK) 
oil-in-soil testing
7 mg/kg (20 g soil : 40 mL solvent, ERACHECK PRO)
54 mg/kg (20 g soil : 40 mL solvent, ERACHECK)
Repeatability ERACHECK PRO   0 mg/L to 9.9 mg/L: ±0.1 mg/L
10 mg/L to 99.9 mg/L: ±0.3 mg/L
100 mg/L to 199.9 mg/L: ±0.5 mg/L
200 mg/L to 2 000 mg/L: ±2.0 mg/L
Repeatability ERACHECK   1.5 mg/L to 9.9 mg/L: ±0.7 mg/L
10 mg/L to 99.9 mg/L: ±1.0 mg/L
100 mg/L to 200 mg/L: ±1.5 mg/L
Typical Extraction Details   12 ml solvent for 250 ml water
40 ml solvent for 20 g soil
Extraction Time   typically 5 minutes (depending on sample conditions)
Interfaces   Built-in PC with Ethernet, USB and RS232 interfaces
Direct LIMS connectivity and output to printer or PC
Optional input by external keyboard, mouse and barcode reader
Remote Control   Remote service via Ethernet interface
PC Software   ERASOFT RCS – remote control Windows® software for multi-instrument remote control, convenient data transfer, viewing spectra and result analysis
Result database   Over 100,000 detailed test reports stored in internal memory
Alarm Tracking   All alarm messages are stored in the database together with the results
Auto Sampler   Optional 10-position auto-sampler – directly attached
Operating Conditions   Temperature range 10°C to 40°C
Humidity up to 90% RH, non condensing
Power requirements   Auto-switching 85-264 V AC, 47-63 Hz, max. 150 W (multi-voltage power supply)
Field application: 12 V DC (vehicle battery) adapter available
Dimensions / Weight   22 x 32 x 28 cm (8.7 x 12.6 x 11 in) / 8 kg (17.6 lb)

Brochure: ERACHECK by Eralytics

Spanish Brochure: ERACHECK by Eralytics

Categories:
, , , , , , , , , , , , , , , , , , .
Tags:
, , , , , , , , , .
SKU: AYA-80-22008-1.